Biobase

Biobase BK-200 Automatic chemistry analyzer

49 sample positions.

ManufacturerBiobase
AreaClinical Chemistry
GroupAnalyzers
Catalogue numberBK-200
Manufacturer websitewww.biobase.com

Product description

49 sample positions.

56 reagent positions.

120 reaction cuvettes.

True 200 tests/hour for single/double reagent tests.

Probe with anti-collision and liquid level detection.


Model: BK-200

  • Analyzer type: random access, benchtop, fully automatic, with separate PC system
  • Throughput: 200 tests per hour for single/double reagent packs
  • Analytical methods: end point, fixed time, kinetic, turbidimetry
  • Certificates: CE, FDA, ISO9001, ISO14001, ISO13485

Sample and reagent unit:

  • Sample positions: 49, including calibration, quality control and emergency samples
  • Reagent positions: 56, constantly refrigerated
  • Sample volume: 2~70 µl (in 0.1 µl steps)
  • Reagent volume: R1&R2: 20~350 µl (in 1 µl steps)
  • Sample and reagent probe: liquid level sensor, anti-collision function
  • Probe cleaning: automatic inside and outside washing
  • Reagent and sample tray cooling: 2°C~8°C (24 h)

Reaction cuvette system:

  • Reaction time: 10 minutes max.
  • Temperature control: 37°C±0.1°C, real-time monitoring
  • Reaction cuvettes: 120 reusable cuvettes (6 mm optical path)
  • Mixing probe: mixing immediately after adding reagent and sample
  • Cuvette washing: automatic 5-step washing
  • Automatic real-time monitoring of cuvette cleanliness

Optical system:

  • Light source: 12V/30W halogen-tungsten lamp, 2000 hours service life
  • Detector: 8 precision photodetectors
  • Wavelengths: 340, 405, 450, 510, 546, 578, 630, 700 nm
  • Linear range: 0.000~3.000 Abs

Calibration and quality control (QC):

  • Calibration: one point, two point, multi point, factor, logit-4P, logit-5P, spline etc.
  • Quality control: QC at any position, with the option to check the reaction curve of every test. Daily/monthly/yearly QC charts, per LOT etc.

Software:

  • STAT function: emergency samples can be added at any time during a run, and emergency patients can be defined when setting up a run.
  • Computer system: Windows 7/8/10, LIS connection (laboratory information system)
  • Data processing: reaction curve and data display functions
  • Result storage capacity (including reaction curves): unlimited

Other:

  • Power supply: AC110/220V±10%, 60/50Hz
  • Power consumption: 300 W
  • Shipping dimensions (W*D*H): 1118*728*1151 mm
  • Gross weight: 130 kg

(1) Reaction carousel

  • 37±0.1°C, real-time monitoring.

(2) Reagent carousel

  • Refrigerated at 2~8°C (24 h).

(3) Mixing probe

  • Teflon coated to prevent cross-contamination.

(4) Sample probe (needle)

  • Liquid level sensor. Anti-collision function. Real-time reagent volume detection.

(5) Cuvette washing probes

  • Independent 5-step washing system.

(6) Software interface

  • User-oriented software, simple and easy to use.