Biobase BK-500 Automatic chemistry analyzer

90 sample positions.

ManufacturerBiobase
AreaClinical Chemistry
GroupAnalyzers
Catalogue numberBK-500
Manufacturer websitewww.biobase.com

Product description

90 sample positions.

120 reaction cuvettes (reusable).

90 reagent positions (R1: 45 & R2: 45), constantly refrigerated.

Automatic 7-step cuvette washing system.

550 tests/hour with STAT function.

One sample probe and two reagent probes, anti-collision function.

Liquid level detection, real-time reagent volume display.


Model: BK-500

  • Analyzer type: random access, discrete, fully automatic
  • Throughput: 550 tests/hour
  • Analytical methods: end point, fixed time, kinetic, two point, turbidimetry
  • Certificates: CE, FDA, ISO9001, ISO14001, ISO13485

Sample unit:

  • Sample positions: 90, including calibration, quality control and emergency samples
  • Sample volume: 2 µl ~ 70 µl (in 0.1 µl steps)
  • Sample probe: liquid level detection
  • Probe washing: automatic inside and outside washing
  • Analysis: up to 45 different tests simultaneously

Reagent carousel:

  • Reagent positions: R1: 45, R2: 45, 90 refrigerated positions in total
  • Reagent volume: 20~350 µl (in 1 µl steps)
  • Temperature control: constant refrigeration at 2~8°C (24 h)
  • Reagent probe: liquid level sensor
  • Probe washing: automatic inside and outside washing

Reaction system:

  • Temperature: 37°C±0.1°C, real-time monitoring
  • Cuvettes: 120 reusable cuvettes (6 mm optical path)
  • Mixing probe: mixing immediately after adding reagent and sample
  • Washing: automatic cuvette washing
  • Reading correction: the system stores the blank value of each cuvette and subtracts it from the reaction value for more accurate results.

Optical system:

  • Light source: 12V/30W halogen-tungsten lamp, 2000 hours service life
  • Detector: 8 precision photodetectors
  • Wavelengths: 340, 405, 450, 510, 546, 578, 630, 700 nm
  • Linear range: 0.000~3.000 Abs

Calibration and quality control (QC):

  • Calibration: one point, two point, multi point, factor, logit-4P, logit-5P, spline
  • Quality control: random QC insertion

Software:

  • STAT function: emergency samples can be added at any time during analysis.
  • Computer system: Windows 7/8/10, LIS connection (laboratory information system)
  • Data processing: reaction curve and data display functions

Other:

  • Power supply: AC110/220V±10%, 60/50Hz
  • Power consumption: 750 W
  • Package dimensions (W*D*H): 1348*928*1399 mm (analyzer), 678*624*548 mm (accessories)
  • Gross weight: 360 kg

(1) Reaction carousel

  • 37±0.1°C, real-time monitoring.

(2) Reagent carousel

  • Refrigerated at 2~8°C (24 h).

(3) Mixing probe

  • Teflon coated to prevent cross-contamination.

(4) Sample probe (needle)

  • Liquid level sensor. Anti-collision function. Real-time reagent volume detection.

(5) Cuvette washing probes

  • Independent 7-step washing system.

(6) Software

  • User-oriented software, simple and easy to use.