Biobase

Biobase BK-200mini Automatic chemistry analyzer

37 sample positions.

ManufacturerBiobase
AreaClinical Chemistry
GroupAnalyzers
Catalogue numberBK-200mini
Manufacturer websitewww.biobase.com

Product description

37 sample positions.

28 reagent positions.

48 reaction cuvettes.

True 200 tests/hour for single/double reagent tests.

Probe with anti-collision and liquid level detection.


Model: BK-200mini

  • Throughput: 200 tests per hour for single/double reagent packs
  • Analytical methods: end point, fixed time, kinetic
  • Certificates: CE, FDA, ISO9001, ISO14001, ISO13485

Sample and reagent system:

  • Sample positions: 37
  • Reagent positions: 28
  • Sample volume: 2~70 µl
  • Reagent volume: 20~350 µl
  • Probe: liquid level sensor, anti-collision function
  • Probe washing: automatic inside and outside washing
  • Reagent and sample tray cooling: 2°C~8°C (24 h), with independent switch

Reaction cuvette system:

  • Temperature control: 37±0.1°C, real-time monitoring
  • Reaction cuvettes: 48 reusable cuvettes, 6 mm optical path
  • Mixing probe: independent mixing
  • Cuvette washing: automatic 3-step washing
  • STAT function (emergency patient): yes

Optical system:

  • Light source: 6V/10W halogen lamp
  • Spectrophotometry: post-spectral
  • Wavelengths: 340, 405, 450, 510, 546, 578, 630, 700 nm
  • Absorbance range: 0~3.0 Abs

Calibration and quality control (QC):

  • Calibration: linear: K factor, one point, two point, multi point; non-linear: Spline, Polygon, Index, Logarithm, Logit-4P, Logit-5P
  • Quality control: real-time QC, Westgard multi rule, cumulative sum check, Twin plot (2D)

Data processing:

  • Software: Windows 7/8/10, 32 or 64-bit
  • LIS: bi-directional, supports HL7 protocol
  • Interface: LAN port
  • Printer: external, with multiple report formats

Other:

  • Power supply: AC110/220V±10%, 60/50Hz
  • Power consumption: 300 W
  • Analyzer dimensions (W*D*H): 625*425*460 mm
  • Package dimensions (W*D*H): 721*531*855 mm
  • Gross weight: 36 kg

(1) Reaction carousel

  • 37±0.1°C, real-time monitoring.

(2) Reagent carousel

  • Refrigerated tray with independent switch.

(3) Mixing probe

  • Teflon coated to prevent cross-contamination.

(4) Sample probe (needle)

  • Liquid level sensor. Anti-collision function. Real-time reagent volume detection.

(5) Cuvette washing probes

  • Independent 3-step washing system.

(6) LAN

  • Access via LAN port.